Welch Allyn AED 10 Automatic External Defibrillators

November 28, 2007 by RecallAware
Filed under: Medicine 

UPDATED - Recall expanded to include additional part numbers.

Audience: Emergency services personnel, risk managers, and consumers
[UPDATE 11/28/2007] Recall expanded to include Welch Allyn AED 10™ Automatic External Defibrillators, manufactured between March 29, 2007 and August 9, 2007, part numbers 970302E, 970308E, 970309E, 970310E, and 970311E.

[Posted 11/05/2007] FDA issued a Class I recall for Welch Allyn AED 10 Automatic External Defibrillators manufactured between March 29, 2007 and August 9, 2007, part numbers 970302E, 970308E, 970310E, and 970311E. These devices are used by emergency or medical personnel, or by others who have taken the appropriate training in cardiopulmonary arrest (heart attack). They analyze an unconscious patient’s heart rhythm and automatically deliver an electrical shock to the heart if needed to restore normal heart rhythm.

There is a possibility that these recalled devices may experience failure or unacceptable delay in analyzing a patient’s ECG resulting in possible failure to deliver the appropriate therapy. The possible failure or delay depends on the location of the defective part that stores an electrical charge on the circuit board. The company plans to replace all affected units and has set up a call center for customers.

Source:Full Article

Comments

Readers of this site appreciate your comments

You must be logged in to post a comment.

References and links to other sites are provided as a courtesy. The recall and safety information provided here is based entirely on manufacturers' and regulatory agencies' press release or other public information. We make no other representations about the accuracy or timeliness of this information